The performance of a fluoroscopic electronic portal imaging device modified for portability.
Affiliation
Christie Hospital NHS Trust, Manchester, UK.Issue Date
1999-10
Metadata
Show full item recordAbstract
Advances in external beam therapy technology have made routine, efficient conformal therapy a reality. With it comes the increasing need for online treatment verification, which is only achievable at present through the use of electronic portal imaging devices (EPIDs). For a large radiotherapy centre, the provision of one EPID per treatment machine proves extremely expensive. This paper details modifications to the design of a commercial fluoroscopic EPID (the SRI-100) to produce a portable system, capable of providing quick, high quality imaging on more than one treatment machine. We describe the necessary hardware and software changes made to the system, as well as the variety of mechanical and quality control checks performed for testing the stability and quality of the imaging. The modified system has been found to be both electronically and mechanically robust, with associated image quality, scaling, distortion and movement similar to other EPIDs in the department. Although the modification was designed specifically to allow for the acquisition of images from multiple treatment machines, it may also enable the operation of the EPID for other uses such as total body irradiation (TBI) treatment verification and a further range of quality control procedures on the linear accelerator itself.Citation
The performance of a fluoroscopic electronic portal imaging device modified for portability. 1999, 72 (862):1000-5 Br J RadiolJournal
The British Journal of RadiologyPubMed ID
10673953Type
ArticleLanguage
enISSN
0007-1285Collections
Related articles
- A literature review of electronic portal imaging for radiotherapy dosimetry.
- Authors: van Elmpt W, McDermott L, Nijsten S, Wendling M, Lambin P, Mijnheer B
- Issue date: 2008 Sep
- Development of a one-stop beam verification system using electronic portal imaging devices for routine quality assurance.
- Authors: Lim S, Ma SY, Jeung TS, Yi BY, Lee SH, Lee S, Cho SJ, Choi J
- Issue date: 2012 Autumn
- Use of an amorphous silicon electronic portal imaging device for multileaf collimator quality control and calibration.
- Authors: Baker SJ, Budgell GJ, MacKay RI
- Issue date: 2005 Apr 7
- Evaluation of electronic portal imaging device for missing tissue compensator design and verification.
- Authors: Roback DM, Gerbi BJ
- Issue date: 1995 Dec
- The use of an electronic portal imaging device for exit dosimetry and quality control measurements.
- Authors: Kirby MC, Williams PC
- Issue date: 1995 Feb 1