Optimization of scanning parameters for multi-slice CT colonography: experiments with synthetic and animal phantoms.
Affiliation
Imaging Science and Biomedical Engineering, University of Manchester, Manchester, UK. k.embleton@man.ac.ukIssue Date
2003-12
Metadata
Show full item recordAbstract
AIM: To determine the optimal collimation, pitch, tube current and reconstruction interval for multi-slice computed tomography (CT) colonography with regard to attaining satisfactory image quality while minimizing patient radiation dose. MATERIALS AND METHODS: Multi-slice CT was performed on plastic, excised pig colon and whole pig phantoms to determine optimal settings. Performance was judged by detection of simulated polyps and statistical measures of the image parameters. Fat and muscle conspicuity was measured from images of dual tube-current prone/supine patient data to derive a measure of tube current effects on tissue contrast. RESULTS: A collimation of 4 x 2.5 mm was sufficient for detection of polyps 4 mm and larger, provided that a reconstruction interval of 1.25 mm was used. A pitch of 1.5 allowed faster scanning and reduced radiation dose without resulting in a loss of important information, i.e. detection of small polyps, when compared with a pitch of 0.75. Tube current and proportional radiation dose could be lowered substantially without deleterious effects on the detection of the air-mucosal interface, however, increased image noise substantially reduced conspicuity of different tissues. CONCLUSION: An optimal image acquisition set-up of 4 x 2.5 mm collimation, reconstruction interval of 1.25 mm, pitch of 1.5 and dual prone/supine scan of 40/100 mA tube current is proposed for our institution for scanning symptomatic patients. Indications are that where CT colonography is used for colonic polyp screening in non-symptomatic patients, a 40 mA tube current could prove satisfactory for both scans.Citation
Optimization of scanning parameters for multi-slice CT colonography: experiments with synthetic and animal phantoms. 2003, 58 (12):955-63 Clin RadiolJournal
Clinical RadiologyDOI
10.1016/S0009-9260(03)00297-6PubMed ID
14654028Type
ArticleLanguage
enISSN
0009-9260ae974a485f413a2113503eed53cd6c53
10.1016/S0009-9260(03)00297-6
Scopus Count
Collections
Related articles
- Multi-detector row CT colonography: effect of collimation, pitch, and orientation on polyp detection in a human colectomy specimen.
- Authors: Taylor SA, Halligan S, Bartram CI, Morgan PR, Talbot IC, Fry N, Saunders BP, Khosraviani K, Atkin W
- Issue date: 2003 Oct
- CT colonography: determination of optimal CT technique using a novel colon phantom.
- Authors: Johnson KT, Johnson CD, Anderson SM, Bruesewitz MR, Mccollough CH
- Issue date: 2004 Mar-Apr
- Optimization of scanning parameters for CT colonography.
- Authors: Power NP, Pryor MD, Martin A, Horrocks J, McLean AM, Reznek RH
- Issue date: 2002 May
- Protocol optimization of multidetector computed tomography colonography using pig colonic phantoms.
- Authors: Won HJ, Choi BI, Kim SH, Kim YI, Youn BJ, Han JK
- Issue date: 2005 Jan
- [Evaluation of different 16-row CT colonography protocols using a porcine model].
- Authors: Luz O, Schäfer J, Dammann F, Vonthein R, Heuschmid M, Claussen CD
- Issue date: 2004 Oct