Performance of a block detector PET scanner in imaging non-pure positron emitters--modelling and experimental validation with 124I.
Affiliation
North Western Medical Physics, Christie Hospital, Manchester, M20 4BX, UK.Issue Date
2004-12-21
Metadata
Show full item recordAbstract
The key performance measures of resolution, count rate, sensitivity and scatter fraction are predicted for a dedicated BGO block detector patient PET scanner (GE Advance) in 2D mode for imaging with the non-pure positron-emitting radionuclides 124I, 55Co, 61Cu, 62Cu, 64Cu and 76Br. Model calculations including parameters of the scanner, decay characteristics of the radionuclides and measured parameters in imaging the pure positron-emitter 18F are used to predict performance according to the National Electrical Manufacturers Association (NEMA) NU 2-1994 criteria. Predictions are tested with measurements made using 124I and show that, in comparison with 18F, resolution degrades by 1.2 mm radially and tangentially throughout the field-of-view (prediction: 1.2 mm), count-rate performance reduces considerably and in close accordance with calculations, sensitivity decreases to 23.4% of that with 18F (prediction: 22.9%) and measured scatter fraction increases from 10.0% to 14.5% (prediction: 14.7%). Model predictions are expected to be equally accurate for other radionuclides and may be extended to similar scanners. Although performance is worse with 124I than 18F, imaging is not precluded in 2D mode. The viability of 124I imaging and performance in a clinical context compared with 18F is illustrated with images of a patient with recurrent thyroid cancer acquired using both [124I]-sodium iodide and [18F]-2-fluoro-2-deoxyglucose.Citation
Performance of a block detector PET scanner in imaging non-pure positron emitters--modelling and experimental validation with 124I. 2004, 49 (24):5505-28 Phys Med BiolJournal
Physics in Medicine and BiologyPubMed ID
15724539Type
ArticleLanguage
enISSN
0031-9155Related articles
- Three-dimensional positron emission tomography imaging with 124I and 86Y.
- Authors: Vandenberghe S
- Issue date: 2006 Mar
- Validation of GATE Monte Carlo simulations of the GE Advance/Discovery LS PET scanners.
- Authors: Schmidtlein CR, Kirov AS, Nehmeh SA, Erdi YE, Humm JL, Amols HI, Bidaut LM, Ganin A, Stearns CW, McDaniel DL, Hamacher KA
- Issue date: 2006 Jan
- Design of a very high-resolution small animal PET scanner using a silicon scatter detector insert.
- Authors: Park SJ, Rogers WL, Clinthorne NH
- Issue date: 2007 Aug 7
- Performance characteristics of a new 3-dimensional continuous-emission and spiral-transmission high-sensitivity and high-resolution PET camera evaluated with the NEMA NU 2-2001 standard.
- Authors: Matsumoto K, Kitamura K, Mizuta T, Tanaka K, Yamamoto S, Sakamoto S, Nakamoto Y, Amano M, Murase K, Senda M
- Issue date: 2006 Jan
- Validation of a Monte Carlo simulation of the Philips Allegro/GEMINI PET systems using GATE.
- Authors: Lamare F, Turzo A, Bizais Y, Le Rest CC, Visvikis D
- Issue date: 2006 Feb 21