Use of an amorphous silicon electronic portal imaging device for multileaf collimator quality control and calibration.
Affiliation
North Western Medical Physics, Christie Hospital NHS Trust, Manchester M20 4BX, UK. sally.baker@physics.cr.man.ac.ukIssue Date
2005-04-07
Metadata
Show full item recordAbstract
Multileaf collimator (MLC) calibration and quality control is a time-consuming procedure typically involving the processing, scanning and analysis of films to measure leaf and collimator positions. Faster and more reliable calibration procedures are required for these tasks, especially with the introduction of intensity modulated radiotherapy which requires more frequent checking and finer positional leaf tolerances than previously. A routine quality control (QC) technique to measure MLC leaf bank gain and offset, as well as minor offsets (individual leaf position relative to a reference leaf), using an amorphous silicon electronic portal imaging device (EPID) has been developed. The technique also tests the calibration of the primary and back-up collimators. A detailed comparison between film and EPID measurements has been performed for six linear accelerators (linacs) equipped with MLC and amorphous silicon EPIDs. Measurements of field size from 4 to 24 cm with the EPID were systematically smaller than film measurements over all field sizes by 0.4 mm for leaves/back-up collimators and by 0.2 mm for conventional collimators. This effect is due to the gain calibration correction applied by the EPID, resulting in a 'flattening' of primary beam profiles. Linac dependent systematic differences of up to 0.5 mm in individual leaf/collimator positions were also found between EPID and film measurements due to the difference between the mechanical and radiation axes of rotation. When corrections for these systematic differences were applied, the residual random differences between EPID and film were 0.23 mm and 0.26 mm (1 standard deviation) for field size and individual leaf/back-up collimator position, respectively. Measured gains (over a distance of 220 mm) always agreed within 0.4 mm with a standard deviation of 0.17 mm. Minor offset measurements gave a mean agreement between EPID and film of 0.01+/-0.10 mm (1 standard deviation) after correction for the tilt of the EPID and small rotational misalignments between leaf banks and the back-up collimators used as a reference straight edge. Reproducibility of EPID measurements was found to be very high, with a standard deviation of <0.05 mm for field size and <0.1 mm for individual leaf/collimator positions for a 10x10 cm2 field. A standard set of QC images (three field sizes defined both by leaves only and collimators only) can be acquired in less than 20 min and analysed in 5 min.Citation
Use of an amorphous silicon electronic portal imaging device for multileaf collimator quality control and calibration. 2005, 50 (7):1377-92 Phys Med BiolJournal
Physics in Medicine and BiologyDOI
10.1088/0031-9155/50/7/003PubMed ID
15798330Type
ArticleLanguage
enISSN
0031-9155ae974a485f413a2113503eed53cd6c53
10.1088/0031-9155/50/7/003
Scopus Count
Collections
Related articles
- Analysis of the measurement precision of an amorphous silicon EPID used for MLC leaf position quality control and the long-term calibration stability of an optically controlled MLC.
- Authors: Budgell GJ, Clarke MF
- Issue date: 2008 Aug 7
- Evaluation of MLC leaf positioning using a scanning liquid ionization chamber EPID.
- Authors: Mohammadi M, Bezak E
- Issue date: 2007 Jan 7
- Evaluation of two methods of predicting MLC leaf positions using EPID measurements.
- Authors: Parent L, Seco J, Evans PM, Dance DR, Fielding A
- Issue date: 2006 Sep
- Use of an amorphous silicon EPID for measuring MLC calibration at varying gantry angle.
- Authors: Clarke MF, Budgell GJ
- Issue date: 2008 Jan 21
- Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapy.
- Authors: Greer PB, Popescu CC
- Issue date: 2003 Jul