    Use of an amorphous silicon EPID for measuring MLC calibration at varying gantry angle.

    • CSV
    • RefMan
    • EndNote
    • BibTex
    • RefWorks
    Authors
    Clarke, Mathew F
    Budgell, Geoff J
    Affiliation
    North Western Medical Physics, Christie Hospital NHS Foundation Trust, Withington, Manchester, M20 4BX, UK.
    Issue Date
    2008-01-21
    
    Metadata
    Show full item record
    Abstract
    Amorphous silicon electronic portal imaging devices (EPIDs) are used to perform routine quality control (QC) checks on the multileaf collimators (MLCs) at this centre. Presently, these checks are performed at gantry angle 0 degrees and are considered to be valid for all other angles. Since therapeutic procedures regularly require the delivery of MLC-defined fields to the patient at a wide range of gantry angles, the accuracy of the QC checks at other gantry angles has been investigated. When the gantry is rotated to angles other than 0 degrees it was found that the apparent pixel size measured using the EPID varies up to a maximum value of 0.0015 mm per pixel due to a sag in the EPID of up to 9.2 mm. A correction factor was determined using two independent methods at a range of gantry angles between 0 degrees and 360 degrees . The EPID was used to measure field sizes (defined by both x-jaws and MLC) at a range of gantry angles and, after this correction had been applied, any residual gravitational sag was studied. It was found that, when fields are defined by the x-jaws and y-back-up jaws, no errors of greater than 0.5 mm were measured and that these errors were no worse when the MLC was used. It was therefore concluded that, provided the correction is applied, measurements of the field size are, in practical terms, unaffected by gantry angle. Experiments were also performed to study how the reproducibility of individual leaves is affected by gantry angle. Measurements of the relative position of each individual leaf (minor offsets) were performed at a range of gantry angles and repeated three times. The position reproducibility was defined by the RMS error in the position of each leaf and this was found to be 0.24 mm and 0.21 mm for the two leaf banks at a gantry angle of 0 degrees . When measurements were performed at a range of gantry angles, these reproducibility values remained within 0.09 mm and 0.11 mm. It was therefore concluded that the calibration of the Elekta MLC is stable at all gantry angles.
    Citation
    Use of an amorphous silicon EPID for measuring MLC calibration at varying gantry angle. 2008, 53 (2):473-85 Phys Med Biol
    Journal
    Physics in Medicine and Biology
    URI
    http://hdl.handle.net/10541/66855
    DOI
    10.1088/0031-9155/53/2/013
    PubMed ID
    18185000
    Type
    Article
    Language
    en
    ISSN
    0031-9155
    ae974a485f413a2113503eed53cd6c53
    10.1088/0031-9155/53/2/013
    Scopus Count
    Collections
    All Christie Publications
    Christie Medical Physics and Engineering Research

    entitlement

    Related articles

    • Use of an amorphous silicon electronic portal imaging device for multileaf collimator quality control and calibration.
    • Authors: Baker SJ, Budgell GJ, MacKay RI
    • Issue date: 2005 Apr 7
    • Evaluation of two methods of predicting MLC leaf positions using EPID measurements.
    • Authors: Parent L, Seco J, Evans PM, Dance DR, Fielding A
    • Issue date: 2006 Sep
    • Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapy.
    • Authors: Greer PB, Popescu CC
    • Issue date: 2003 Jul
    • Detection and correction for EPID and gantry sag during arc delivery using cine EPID imaging.
    • Authors: Rowshanfarzad P, Sabet M, O'Connor DJ, McCowan PM, McCurdy BM, Greer PB
    • Issue date: 2012 Feb
    • Evaluation of MLC leaf positioning using a scanning liquid ionization chamber EPID.
    • Authors: Mohammadi M, Bezak E
    • Issue date: 2007 Jan 7
    DSpace software (copyright © 2002 - 2025)  DuraSpace
    Quick Guide | Contact Us
    Open Repository is a service operated by 
    Atmire NV
     

    Export search results

    The export option will allow you to export the current search results of the entered query to a file. Different formats are available for download. To export the items, click on the button corresponding with the preferred download format.

    By default, clicking on the export buttons will result in a download of the allowed maximum amount of items.

    To select a subset of the search results, click "Selective Export" button and make a selection of the items you want to export. The amount of items that can be exported at once is similarly restricted as the full export.

    After making a selection, click one of the export format buttons. The amount of items that will be exported is indicated in the bubble next to export format.