Show simple item record

dc.contributor.authorTran-Gia, Jen
dc.contributor.authorDenis-Bacelar, AMen
dc.contributor.authorFerreira, KMen
dc.contributor.authorRobinson, APen
dc.contributor.authorBobin, Cen
dc.contributor.authorBonney, LMen
dc.contributor.authorCalvert, Nicolasen
dc.contributor.authorCollins, SMen
dc.contributor.authorFenwick, AJen
dc.contributor.authorFinocchiaro, Den
dc.contributor.authorFioroni, Fen
dc.contributor.authorGiannopoulou, Ken
dc.contributor.authorGrassi, Een
dc.contributor.authorHeetun, Wen
dc.contributor.authorJewitt, SJen
dc.contributor.authorKotzasarlidou, Men
dc.contributor.authorLjungberg, Men
dc.contributor.authorLourenço, Ven
dc.contributor.authorMcGowan, DRen
dc.contributor.authorMewburn-Crook, Jen
dc.contributor.authorSabot, Ben
dc.contributor.authorScuffham, Jen
dc.contributor.authorSjögreen Gleisner, Ken
dc.contributor.authorSolc, Jen
dc.contributor.authorThiam, Cen
dc.contributor.authorTipping, Jen
dc.contributor.authorWevrett, Jen
dc.contributor.authorLassmann, Men
dc.date.accessioned2024-01-29T13:19:45Z
dc.date.available2024-01-29T13:19:45Z
dc.date.issued2023en
dc.identifier.citationTran-Gia J, Denis-Bacelar AM, Ferreira KM, Robinson AP, Bobin C, Bonney LM, et al. On the use of solid (133)Ba sources as surrogate for liquid (131)I in SPECT/CT calibration: a European multi-centre evaluation. EJNMMI physics. 2023 Nov 23;10(1):73. PubMed PMID: 37993667. Pubmed Central PMCID: PMC10665282. Epub 2023/11/23. eng.en
dc.identifier.pmid37993667en
dc.identifier.doi10.1186/s40658-023-00582-3en
dc.identifier.urihttp://hdl.handle.net/10541/626835
dc.description.abstractINTRODUCTION: Commissioning, calibration, and quality control procedures for nuclear medicine imaging systems are typically performed using hollow containers filled with radionuclide solutions. This leads to multiple sources of uncertainty, many of which can be overcome by using traceable, sealed, long-lived surrogate sources containing a radionuclide of comparable energies and emission probabilities. This study presents the results of a quantitative SPECT/CT imaging comparison exercise performed within the MRTDosimetry consortium to assess the feasibility of using (133)Ba as a surrogate for (131)I imaging. MATERIALS AND METHODS: Two sets of four traceable (133)Ba sources were produced at two National Metrology Institutes and encapsulated in 3D-printed cylinders (volume range 1.68-107.4 mL). Corresponding hollow cylinders to be filled with liquid (131)I and a mounting baseplate for repeatable positioning within a Jaszczak phantom were also produced. A quantitative SPECT/CT imaging comparison exercise was conducted between seven members of the consortium (eight SPECT/CT systems from two major vendors) based on a standardised protocol. Each site had to perform three measurements with the two sets of (133)Ba sources and liquid (131)I. RESULTS: As anticipated, the (131)I pseudo-image calibration factors (cps/MBq) were higher than those for (133)Ba for all reconstructions and systems. A site-specific cross-calibration reduced the performance differences between both radionuclides with respect to a cross-calibration based on the ratio of emission probabilities from a median of 12-1.5%. The site-specific cross-calibration method also showed agreement between (133)Ba and (131)I for all cylinder volumes, which highlights the potential use of (133)Ba sources to calculate recovery coefficients for partial volume correction. CONCLUSION: This comparison exercise demonstrated that traceable solid (133)Ba sources can be used as surrogate for liquid (131)I imaging. The use of solid surrogate sources could solve the radiation protection problem inherent in the preparation of phantoms with (131)I liquid activity solutions as well as reduce the measurement uncertainties in the activity. This is particularly relevant for stability measurements, which have to be carried out at regular intervals.en
dc.language.isoenen
dc.relation.urlhttps://dx.doi.org/10.1186/s40658-023-00582-3en
dc.titleOn the use of solid (133)Ba sources as surrogate for liquid (131)I in SPECT/CT calibration: a European multi-centre evaluationen
dc.typeArticleen
dc.contributor.departmentDepartment of Nuclear Medicine, University Hospital Würzburg, Oberdürrbacher Str. 6, 97080, Würzburg, Germany. Tran_J@ukw.de. Christie Medical Physics and Engineering (CMPE), The Christie NHS Foundation Trust, Manchester, UK.en
dc.identifier.journalEJNMMI Physicsen
dc.description.noteen]
refterms.dateFOA2024-02-01T13:20:45Z


Files in this item

Thumbnail
Name:
On the use of solid....pdf
Size:
1.313Mb
Format:
PDF
Description:
Found with Open Access Button

This item appears in the following Collection(s)

Show simple item record