A portable primary-standard level graphite calorimeter for absolute dosimetry in clinical pencil beam scanning proton beams
Authors
Lourenco, A.Lee, N.
Charlwood, Frances
Lambert, J.
Vera, J. A.
Hussein, M.
Shipley, D. R.
Romano, F.
Lowe, Matthew
Clarke, Matthew
Lorentini, S.
Mazal, A.
Pettingell, J.
Palmans, H.
Thomas, R. A. S.
Affiliation
Medical Radiation Science Group, National Physical Laboratory, Hampton Rd, Teddington, TW11 0LW,Issue Date
2023
Metadata
Show full item recordAbstract
To report the use of a portable primary standard level graphite calorimeter for direct dose determination in clinical pencil beam scanning proton beams, which forms part of the recommendations of the proposed Institute of Physics and Engineering in Medicine (IPEM) Code of Practice (CoP) for proton therapy dosimetry.
Approach: The Primary Standard Proton Calorimeter (PSPC) was developed at the National Physical Laboratory (NPL) and measurements were performed at four clinical proton therapy facilities that use pencil beam scanning for beam delivery. Correction factors for the presence of impurities and vacuum gaps were calculated and applied, as well as dose conversion factors to obtain dose-to-water. Measurements were performed in the middle of 10×10×10 cm3 homogeneous dose volumes, centred at 10.0, 15.0 and 25.0 g∙cm-2 depth in water. The absorbed dose to water determined with the calorimeter was compared to the dose obtained using PTW Roos-type ionisation chambers calibrated in terms of absorbed dose to water in 60Co applying the recommendations in the IAEA TRS-398 CoP.
Main results: The relative dose difference between the two protocols varied between 0.4% and 2.1% depending on the facility. The reported overall uncertainty in the determination of absorbed dose-to-water using the calorimeter is 0.9% (k=1), which corresponds to a significant reduction of uncertainty in comparison with the TRS-398 CoP (currently with an uncertainty equal or larger than 2.0% (k=1) for proton beams).
Significance: The establishment of a purpose-built primary standard and associated CoP will considerably reduce the uncertainty of the absorbed dose-to-water determination and ensure improved accuracy and consistency in the dose delivered to patients treated with proton therapy and bring proton reference dosimetry uncertainty in line with megavoltage photon radiotherapy.Citation
Lourenco A, Lee N, Charlwood F, Lambert J, Vera JA, Hussein M, et al. A portable primary-standard level graphite calorimeter for absolute dosimetry in clinical pencil beam scanning proton beams. Physics in medicine and biology. 2023 Jul 6. PubMed PMID: 37414003. Epub 2023/07/07. eng.Journal
Physics in Medicine and BiologyDOI
10.1088/1361-6560/ace50fPubMed ID
37414003Additional Links
https://dx.doi.org/10.1088/1361-6560/ace50fType
ArticleLanguage
enae974a485f413a2113503eed53cd6c53
10.1088/1361-6560/ace50f
Scopus Count
Collections
Related articles
- Application of a portable primary standard level graphite calorimeter for absolute dosimetry in a clinical low-energy passively scattered proton beam.
- Authors: Lourenço A, Lee N, Shipley D, Romano F, Kacperek A, Duane S, Cashmore M, Bass G, Palmans H, Thomas R
- Issue date: 2022 Nov 18
- Ultrahigh dose rate pencil beam scanning proton dosimetry using ion chambers and a calorimeter in support of first in-human FLASH clinical trial.
- Authors: Lee E, Lourenço AM, Speth J, Lee N, Subiel A, Romano F, Thomas R, Amos RA, Zhang Y, Xiao Z, Mascia A
- Issue date: 2022 Sep
- A small-body portable graphite calorimeter for dosimetry in low-energy clinical proton beams.
- Authors: Palmans H, Thomas R, Simon M, Duane S, Kacperek A, DuSautoy A, Verhaegen F
- Issue date: 2004 Aug 21
- Experimental comparison of cylindrical and plane parallel ionization chambers for reference dosimetry in continuous and pulsed scanned proton beams.
- Authors: Vilches-Freixas G, Bosmans G, Douralis A, Martens J, Meijers A, Rinaldi I, Salvo K, Thomas R, Palmans H, Lourenço A
- Issue date: 2024 May 8
- Direct absorbed dose to water determination based on water calorimetry in scanning proton beam delivery.
- Authors: Sarfehnia A, Clasie B, Chung E, Lu HM, Flanz J, Cascio E, Engelsman M, Paganetti H, Seuntjens J
- Issue date: 2010 Jul