    Evaluation of GATE-RTion (GATE/Geant4) Monte Carlo simulation settings for proton pencil beam scanning quality assurance

    • CSV
    • RefMan
    • EndNote
    • BibTex
    • RefWorks
    Thumbnail
    Name:
    32967037.pdf
    Size:
    7.168Mb
    Format:
    PDF
    Description:
    From UNPAYWALL
    Download
    Authors
    Winterhalter, Carla
    Taylor, Michael J
    Boersma, D.
    Elia, A.
    Guatelli, S.
    Mackay, Ranald I
    Kirkby, Karen J
    Maigne, L.
    Ivanchenko, V.
    Resch, A. F.
    Sarrut, D.
    Sitch, Peter
    Vidal, M.
    Grevillot, L.
    Aitkenhead, Adam H
    Show allShow less
    Affiliation
    Division of Cancer Sciences, University of Manchester, Manchester, M13 9PL, UK. T
    Issue Date
    2020
    
    Metadata
    Show full item record
    Abstract
    Purpose: Geant4 is a multi-purpose Monte Carlo simulation tool for modeling particle transport in matter. It provides a wide range of settings, which the user may optimize for their specific application. This study investigates GATE/Geant4 parameter settings for proton pencil beam scanning therapy. Methods: GATE8.1/Geant4.10.3.p03 (matching the versions used in GATE-RTion1.0) simulations were performed with a set of prebuilt Geant4 physics lists (QGSP_BIC, QGSP_BIC_EMY, QGSP_BIC_EMZ, QGSP_BIC_HP_EMZ), using 0.1mm-10mm as production cuts on secondary particles (electrons, photons, positrons) and varying the maximum step size of protons (0.1mm, 1mm, none). The results of the simulations were compared to measurement data taken during clinical patient specific quality assurance at The Christie NHS Foundation Trust pencil beam scanning proton therapy facility. Additionally, the influence of simulation settings was quantified in a realistic patient anatomy based on computer tomography (CT) scans. Results: When comparing the different physics lists, only the results (ranges in water) obtained with QGSP_BIC (G4EMStandardPhysics_Option0) depend on the maximum step size. There is clinically negligible difference in the target region when using High Precision neutron models (HP) for dose calculations. The EMZ electromagnetic constructor provides a closer agreement (within 0.35 mm) to measured beam sizes in air, but yields up to 20% longer execution times compared to the EMY electromagnetic constructor (maximum beam size difference 0.79 mm). The impact of this on patient-specific quality assurance simulations is clinically negligible, with a 97% average 2%/2 mm gamma pass rate for both physics lists. However, when considering the CT-based patient model, dose deviations up to 2.4% are observed. Production cuts do not substantially influence dosimetric results in solid water, but lead to dose differences of up to 4.1% in the patient CT. Small (compared to voxel size) production cuts increase execution times by factors of 5 (solid water) and 2 (patient CT). Conclusions: Taking both efficiency and dose accuracy into account and considering voxel sizes with 2 mm linear size, the authors recommend the following Geant4 settings to simulate patient specific quality assurance measurements: No step limiter on proton tracks; production cuts of 1 mm for electrons, photons and positrons (in the phantom and range-shifter) and 10 mm (world); best agreement to measurement data was found for QGSP_BIC_EMZ reference physics list at the cost of 20% increased execution times compared to QGSP_BIC_EMY. For simulations considering the patient CT model, the following settings are recommended: No step limiter on proton tracks; production cuts of 1 mm for electrons, photons and positrons (phantom/range-shifter) and 10 mm (world) if the goal is to achieve sufficient dosimetric accuracy to ensure that a plan is clinically safe; or 0.1 mm (phantom/range-shifter) and 1 mm (world) if higher dosimetric accuracy is needed (increasing execution times by a factor of 2); most accurate results expected for QGSP_BIC_EMZ reference physics list, at the cost of 10-20% increased execution times compared to QGSP_BIC_EMY.
    Citation
    Winterhalter C, Taylor M, Boersma D, Elia A, Guatelli S, Mackay R, et al. Evaluation of GATE-RTion (GATE/Geant4) Monte Carlo simulation settings for proton pencil beam scanning quality assurance. Med Phys. 2020.
    Journal
    Medical Physics
    URI
    http://hdl.handle.net/10541/623450
    DOI
    10.1002/mp.14481
    PubMed ID
    32967037
    Additional Links
    https://dx.doi.org/10.1002/mp.14481
    Type
    Article
    Language
    en
    ae974a485f413a2113503eed53cd6c53
    10.1002/mp.14481
    Scopus Count
    Collections
    All Christie Publications

    entitlement

    Related articles

    • Experimental validation of proton physics models of Geant4 for calculating stopping power ratio.
    • Authors: Liu R, Zhao X, Medrano M
    • Issue date: 2022 Jun 28
    • Optimization of Monte Carlo particle transport parameters and validation of a novel high throughput experimental setup to measure the biological effects of particle beams.
    • Authors: Patel D, Bronk L, Guan F, Peeler CR, Brons S, Dokic I, Abdollahi A, Rittmüller C, Jäkel O, Grosshans D, Mohan R, Titt U
    • Issue date: 2017 Nov
    • TOPAS/Geant4 configuration for ionization chamber calculations in proton beams.
    • Authors: Wulff J, Baumann KS, Verbeek N, Bäumer C, Timmermann B, Zink K
    • Issue date: 2018 Jun 7
    • GATE MODELING OF LATERAL SCATTERING OF PROTON PENCIL BEAMS.
    • Authors: Zarifi S, Ahangari HT, Jia SB, Tajik-Mansoury MA, Najafzadeh M
    • Issue date: 2020 Jul 7
    • Benchmarking a GATE/Geant4 Monte Carlo model for proton beams in magnetic fields.
    • Authors: Padilla-Cabal F, Alejandro Fragoso J, Franz Resch A, Georg D, Fuchs H
    • Issue date: 2020 Jan
    DSpace software (copyright © 2002 - 2023)  DuraSpace
    Quick Guide | Contact Us
    Open Repository is a service operated by 
    Atmire NV
     

    Export search results

    The export option will allow you to export the current search results of the entered query to a file. Different formats are available for download. To export the items, click on the button corresponding with the preferred download format.

    By default, clicking on the export buttons will result in a download of the allowed maximum amount of items.

    To select a subset of the search results, click "Selective Export" button and make a selection of the items you want to export. The amount of items that can be exported at once is similarly restricted as the full export.

    After making a selection, click one of the export format buttons. The amount of items that will be exported is indicated in the bubble next to export format.