Characterizing local dose perturbations due to gas cavities in magnetic resonance guided radiotherapy
Authors
Shortall, JVasquez Osorio, Eliana
Chuter, Robert
Green, Andrew
McWilliam, Alan
Kirkby, Karen J
Mackay, Ranald I
van Herk, Marcel
Affiliation
The University of Manchester, Division of Cancer Sciences, Faculty of Biology, Medicine and Health, ManchesterIssue Date
2020
Metadata
Show full item recordAbstract
PURPOSE: Due to differences in attenuation and the Electron Return Effect (ERE), the presence of gas can increase the risk of toxicity in Organs At risk (OAR) during Magnetic Resonance guided Radiotherapy (MRgRT). Current adaptive MRgRT workflows using density overrides negate gas from the dose calculation, meaning that the effects of ERE around gas are not taken into account. In order to achieve an accurate adaptive MRgRT treatment, we should be able to quickly evaluate whether gas present during treatment causes dose constraint violation during an MRgRT fraction. We propose an analytic method for predicting dose perturbations caused by air cavities in OARs during MRgRT. METHOD: Ten virtual water phantoms were created: nine containing a centrally located spherical air cavity and a reference phantom without an air cavity. Monte Carlo dose calculations were produced to irradiate the phantoms with a single 7 MV photon beam under the influence of a 1.5 T transverse magnetic field (Monaco 5.19.02 Treatment Panning System (TPS) (Elekta AB, Stockholm, Sweden)). Dose distributions of the phantoms with and without air cavities were compared. We used a spherical coordinate system originating in the centre of the cavity to sample the dose distributions and calculate the dose perturbation as a result of the presence of each air cavity, ∆D%(θ,Φ)calc . . Dose effects due to ERE and differences in attenuation due to density changes were considered separately. Least squared analysis was used to fit the calculated dose perturbations to mathematical functions. Effects due to ERE were fit to a modulated sinusoidal function and those due to attenuation differences were fit to a 2D Gaussian function. We used the fits to derive a single equation describing dose perturbations around spherical air cavities as a function of angles, θ, Φ, distance from cavity surface, d, and cavity radius, r. We measured the fitting error by calculating the Residual Error (RE); the difference between the calculated and fitted dose perturbation. RESULTS: Both ERE and differences in attenuation contribute towards the total dose effects of air cavities in MRgRT. Whereas ERE dominates close to the surface of the cavities, attenuation effects dominate at distances >0.5 cm from the cavities. We showed that dose effects around a spherical air cavity (≤1 cm from the surface) due to ERE fit a modulated sinusoidal function with mean(RE) ≤-1.4E-5 % and root mean square error (rms)(RE) ≤ 4.1 %. Effects due to attenuation differences fit a Gaussian function with mean(RE) ≤0.7 % and rms(RE) ≤1.8 %. Our general equation, which we verified using multiple sizes of spherical and cylindrical air cavity, fits Monte Carlo simulated data with mean(RE) ≤±0.9 % and rms(RE) ≤6.9 %. CONCLUSION: We show that local dose perturbations around unplanned spherical air cavities during MRgRT can be well characterized analytically. We present an equation that can be incorporated into the clinical workflow to allow for fast evaluation of dose effects of unplanned gas. We also envision this method contributing to the clinical implementation of real time Adaptive Radiotherapy (ART) for MRgRT using MRI planning.Citation
Shortall J, Vasquez Osorio E, Chuter R, Green A, McWilliam A, Kirkby K, et al. Characterizing local dose perturbations due to gas cavities in Magnetic Resonance guided Radiotherapy. Med Phys. 2020.Journal
Medical PhysicsDOI
10.1002/mp.14120PubMed ID
32144781Additional Links
https://dx.doi.org/10.1002/mp.14120Type
ArticleLanguage
enae974a485f413a2113503eed53cd6c53
10.1002/mp.14120
Scopus Count
Collections
Related articles
- Assessing localized dosimetric effects due to unplanned gas cavities during pelvic MR-guided radiotherapy using Monte Carlo simulations.
- Authors: Shortall J, Vasquez Osorio E, Chuter R, McWilliam A, Choudhury A, Kirkby K, Mackay R, van Herk M
- Issue date: 2019 Dec
- Experimental verification the electron return effect around spherical air cavities for the MR-Linac using Monte Carlo calculation.
- Authors: Shortall J, Vasquez Osorio E, Aitkenhead A, Berresford J, Agnew J, Budgell G, Chuter R, McWilliam A, Kirkby K, Mackay R, van Herk M
- Issue date: 2020 Jun
- A 1.5 T transverse magnetic field in radiotherapy of rectal cancer: Impact on the dose distribution.
- Authors: Uilkema S, van der Heide U, Sonke JJ, Moreau M, van Triest B, Nijkamp J
- Issue date: 2015 Dec
- Compensating for the impact of non-stationary spherical air cavities on IMRT dose delivery in transverse magnetic fields.
- Authors: Bol GH, Lagendijk JJ, Raaymakers BW
- Issue date: 2015 Jan 21
- Development and evaluation of a GEANT4-based Monte Carlo Model of a 0.35 T MR-guided radiation therapy (MRgRT) linear accelerator.
- Authors: Khan AU, Simiele EA, Lotey R, DeWerd LA, Yadav P
- Issue date: 2021 Apr