Quantitative evaluation of 4D cone beam CT scans with reduced scan time in lung cancer patients
Authors
Bryce-Atkinson, AMarchant, Thomas E
Rodgers, John
Budgell, Geoff J
McWilliam, Alan
Faivre-Finn, Corinne
Whitfield, Gillian A
van Herk, Marcel
Affiliation
Division of Cancer Sciences, School of Medical Sciences, Faculty of Biology, Medicine and Health, The University of Manchester, UK.Issue Date
2019
Metadata
Show full item recordAbstract
PURPOSE: Image guided radiotherapy (IGRT) based on respiration correlated cone-beam CT (4D-CBCT) provides accurate tumour localisation in lung cancer patients by taking into account respiratory motion when deriving setup correction. However, 4D-CBCT scan times are typically longer than for acquisition of 3D-CBCT scans, e.g. 4?min. This work aims to quantitatively evaluate the effect of reduced scan times on 4D-CBCT image quality and registration accuracy in lung cancer patients. METHODS AND MATERIALS: Scan times down to 1?min were simulated by retaining only projection images corresponding to every second, third or fourth respiratory cycle in forty-four 4D-CBCTs from 15 lung cancer patients. In addition twenty 2-minute scans were acquired for 12 lung cancer patients. Image quality was quantified by assessing registration accuracy in the shorter scan times, comparing to the 4-minute scan registration result where available as reference. RESULTS: Use of 2-minute scans had little impact on registration accuracy or ability to detect tumour motion: automatic registration accuracy was within 2?mm in 6/8 scans analysed with 2-minute acquisitions, and 96.6% of registration discrepancies were within 2?mm for the simulated scans. When the scan time simulated was below 2?min, automatic registration results still agreed within 2?mm for 84.7% of scans, however visual image quality was considerably degraded. CONCLUSION: A 4D-CBCT acquisition time of 2?min produces scans of sufficient image quality for IGRT in most lung cancer patients, as demonstrated quantitatively by assessing the impact on automatic registration accuracy in simulated and real acquisitions.Citation
Bryce-Atkinson A, Marchant T, Rodgers J, Budgell G, McWilliam A, Faivre-Finn C, et al. Quantitative evaluation of 4D cone beam CT scans with reduced scan time in lung cancer patients. Radiother Oncol. 2019; 136:64-70.Journal
Radiotherapy and OncologyDOI
10.1016/j.radonc.2019.03.027PubMed ID
31015131Additional Links
https://dx.doi.org/10.1016/j.radonc.2019.03.027Type
ArticleLanguage
enae974a485f413a2113503eed53cd6c53
10.1016/j.radonc.2019.03.027
Scopus Count
Collections
Related articles
- Investigation of respiration induced intra- and inter-fractional tumour motion using a standard Cone Beam CT.
- Authors: Gottlieb KL, Hansen CR, Hansen O, Westberg J, Brink C
- Issue date: 2010 Oct
- Comparative study of respiratory motion correction techniques in cone-beam computed tomography.
- Authors: Rit S, Nijkamp J, van Herk M, Sonke JJ
- Issue date: 2011 Sep
- 4D liver tumor localization using cone-beam projections and a biomechanical model.
- Authors: Zhang Y, Folkert MR, Li B, Huang X, Meyer JJ, Chiu T, Lee P, Tehrani JN, Cai J, Parsons D, Jia X, Wang J
- Issue date: 2019 Apr
- Estimating 4D-CBCT from prior information and extremely limited angle projections using structural PCA and weighted free-form deformation for lung radiotherapy.
- Authors: Harris W, Zhang Y, Yin FF, Ren L
- Issue date: 2017 Mar
- Accuracy and inter-observer variability of 3D versus 4D cone-beam CT based image-guidance in SBRT for lung tumors.
- Authors: Sweeney RA, Seubert B, Stark S, Homann V, Müller G, Flentje M, Guckenberger M
- Issue date: 2012 Jun 8