Improving molecular radiotherapy dosimetry using anthropomorphic calibration
Name:
1-s2.0-S1120179719300134-main.pdf
Size:
624.1Kb
Format:
PDF
Description:
Full text, Open Access article
Authors
Price, ERobinson, P
Cullen, M
Tipping, Jill
Calvert, Nicholas
Hamilton, David
Oldfield, C
Page, Emma
Pietras, B
Smith, A
Affiliation
Division of Cancer Sciences, Faculty of Biology, Medicine & Health, University of Manchester, Manchester Academic Health Sciences Centre, ManchesterIssue Date
2019
Metadata
Show full item recordAbstract
The optimised delivery of Molecular Radiotherapy requires individualised calculation of absorbed dose to both targeted lesions and neighbouring healthy tissue. To achieve this, accurate quantification of the activity distribution in the patient by external detection is vital. METHODS: This work extends specific anatomy-related calibration to true organ shapes. A set of patient-specific 3D printed organ inserts based on a diagnostic CT scan was produced, comprising the liver, spleen and both kidneys. The inserts were used to calculate patient-specific calibration factors for 177Lu. These calibration factors were compared with previously reported calibration factors for corresponding organ models based on the Cristy and Eckerman phantom series and for a comparably sized sphere. Monte Carlo calculations of the patient-specific radiation dose were performed for comparison with current clinical dosimetry methods for these data. RESULTS: Patient-specific calibration factors are shown to be dependent on the volume, shape and position of the organ containing activity with a corresponding impact on the calculation of the dose to the patient. The impact of organ morphology on calculated dose is reduced when the dominant contributor to dose is beta particles. This is due to the small range of beta particles in tissue. Overestimations of recovered activity and hence dose of up to 135% are observed. CONCLUSION: For accurate quantification to be performed calibration factors accounting for organ size, shape and position must be used. Such quantification is vital if accurate, patient-specific dosimetry is to be achieved.Citation
Price E, Robinson A, Cullen D, Tipping J, Calvert N, Hamilton D, et al. Improving molecular radiotherapy dosimetry using anthropomorphic calibration. Phys Med. 2019 Feb;58:40-6.Journal
Physica MedicaDOI
10.1016/j.ejmp.2019.01.013PubMed ID
30824148Additional Links
https://dx.doi.org/10.1016/j.ejmp.2019.01.013Type
ArticleLanguage
enae974a485f413a2113503eed53cd6c53
10.1016/j.ejmp.2019.01.013
Scopus Count
Collections
Related articles
- Organ-specific SPECT activity calibration using 3D printed phantoms for molecular radiotherapy dosimetry.
- Authors: Robinson AP, Tipping J, Cullen DM, Hamilton D, Brown R, Flynn A, Oldfield C, Page E, Price E, Smith A, Snee R
- Issue date: 2016 Dec
- Toward automated and personalized organ dose determination in CT examinations - A comparison of two tissue characterization models for Monte Carlo organ dose calculation with a Therapy Planning System.
- Authors: Källman HE, Traneus E, Ahnesjö A
- Issue date: 2019 Feb
- Design and Fabrication of Kidney Phantoms for Internal Radiation Dosimetry Using 3D Printing Technology.
- Authors: Tran-Gia J, Schlögl S, Lassmann M
- Issue date: 2016 Dec
- Validation of Monte Carlo (131) I radiopharmaceutical dosimetry workflow using a 3D-printed anthropomorphic head and neck phantom.
- Authors: Adam DP, Grudzinski JJ, Bormett I, Cox BL, Marsh IR, Bradshaw TJ, Harari PM, Bednarz BP
- Issue date: 2022 Aug
- On the use of Monte Carlo-derived dosimetric data in the estimation of patient dose from CT examinations.
- Authors: Perisinakis K, Tzedakis A, Damilakis J
- Issue date: 2008 May