Affiliation
Department of Radiation Oncology, The Netherlands Cancer Institute, AmsterdamIssue Date
2016-07
Metadata
Show full item recordAbstract
Delivery errors during radiotherapy may lead to medical harm and reduced life expectancy for patients. Such serious incidents can be avoided by performing dose verification online, i.e., while the patient is being irradiated, creating the possibility of halting the linac in case of a large overdosage or underdosage. The offline EPID-based 3D in vivo dosimetry system clinically employed at our institute is in principle suited for online treatment verification, provided the system is able to complete 3D dose reconstruction and verification within 420 ms, the present acquisition time of a single EPID frame. It is the aim of this study to show that our EPID-based dosimetry system can be made fast enough to achieve online 3D in vivo dose verification.Citation
Online 3D EPID-based dose verification: Proof of concept. 2016, 43 (7):3969 Med PhysJournal
Medical PhysicsDOI
10.1118/1.4952729PubMed ID
27370115Type
ArticleLanguage
enISSN
0094-2405ae974a485f413a2113503eed53cd6c53
10.1118/1.4952729
Scopus Count
Collections
Related articles
- Frame average optimization of cine-mode EPID images used for routine clinical in vivo patient dose verification of VMAT deliveries.
- Authors: McCowan PM, McCurdy BM
- Issue date: 2016 Jan
- Validation of a method for in vivo 3D dose reconstruction for IMRT and VMAT treatments using on-treatment EPID images and a model-based forward-calculation algorithm.
- Authors: Van Uytven E, Van Beek T, McCowan PM, Chytyk-Praznik K, Greer PB, McCurdy BM
- Issue date: 2015 Dec
- Clinical experience with EPID dosimetry for prostate IMRT pre-treatment dose verification.
- Authors: McDermott LN, Wendling M, van Asselen B, Stroom J, Sonke JJ, van Herk M, Mijnheer BJ
- Issue date: 2006 Oct
- An in vivo dose verification method for SBRT-VMAT delivery using the EPID.
- Authors: McCowan PM, Van Uytven E, Van Beek T, Asuni G, McCurdy BM
- Issue date: 2015 Dec
- In aqua vivo EPID dosimetry.
- Authors: Wendling M, McDermott LN, Mans A, Olaciregui-Ruiz Í, Pecharromán-Gallego R, Sonke JJ, Stroom J, van Herk M, Mijnheer BJ
- Issue date: 2012 Jan