Show simple item record

dc.contributor.authorDewhurst, Joseph M
dc.contributor.authorLowe, Matthew
dc.contributor.authorHardy, Mark J
dc.contributor.authorBoylan, Christopher J
dc.contributor.authorWhitehurst, Philip
dc.contributor.authorRowbottom, Carl G
dc.date.accessioned2015-07-21T10:31:12Zen
dc.date.available2015-07-21T10:31:12Zen
dc.date.issued2015en
dc.identifier.citationAutoLock: a semiautomated system for radiotherapy treatment plan quality control. 2015, 16 (3):5396 J Appl Clin Med Physen
dc.identifier.issn1526-9914en
dc.identifier.pmid26103498en
dc.identifier.urihttp://hdl.handle.net/10541/560767en
dc.description.abstractA semiautomated system for radiotherapy treatment plan quality control (QC), named AutoLock, is presented. AutoLock is designed to augment treatment plan QC by automatically checking aspects of treatment plans that are well suited to computational evaluation, whilst summarizing more subjective aspects in the form of a checklist. The treatment plan must pass all automated checks and all checklist items must be acknowledged by the planner as correct before the plan is finalized. Thus AutoLock uniquely integrates automated treatment plan QC, an electronic checklist, and plan finalization. In addition to reducing the potential for the propagation of errors, the integration of AutoLock into the plan finalization workflow has improved efficiency at our center. Detailed audit data are presented, demonstrating that the treatment plan QC rejection rate fell by around a third following the clinical introduction of AutoLock.
dc.language.isoenen
dc.rightsArchived with thanks to Journal of applied clinical medical physics / American College of Medical Physicsen
dc.titleAutoLock: a semiautomated system for radiotherapy treatment plan quality control.en
dc.typeArticleen
dc.contributor.departmentChristie Medical Physics & Engineering The Christie NHS Foundation Trusten
dc.identifier.journalJournal of applied clinical medical physics / American College of Medical Physicsen
html.description.abstractA semiautomated system for radiotherapy treatment plan quality control (QC), named AutoLock, is presented. AutoLock is designed to augment treatment plan QC by automatically checking aspects of treatment plans that are well suited to computational evaluation, whilst summarizing more subjective aspects in the form of a checklist. The treatment plan must pass all automated checks and all checklist items must be acknowledged by the planner as correct before the plan is finalized. Thus AutoLock uniquely integrates automated treatment plan QC, an electronic checklist, and plan finalization. In addition to reducing the potential for the propagation of errors, the integration of AutoLock into the plan finalization workflow has improved efficiency at our center. Detailed audit data are presented, demonstrating that the treatment plan QC rejection rate fell by around a third following the clinical introduction of AutoLock.


This item appears in the following Collection(s)

Show simple item record