Dose distribution for dental cone beam CT and its implication for defining a dose index.
dc.contributor.author | Pauwels, R | |
dc.contributor.author | Theodorakou, C | |
dc.contributor.author | Walker, Anne | |
dc.contributor.author | Bosmans, H | |
dc.contributor.author | Jacobs, R | |
dc.contributor.author | Horner, K | |
dc.contributor.author | Bogaerts, R | |
dc.date.accessioned | 2012-12-28T15:54:39Z | |
dc.date.available | 2012-12-28T15:54:39Z | |
dc.date.issued | 2012-10 | |
dc.identifier.citation | Dose distribution for dental cone beam CT and its implication for defining a dose index. 2012, 41 (7):583-93 Dentomaxillofac Radiol | en_GB |
dc.identifier.issn | 0250-832X | |
dc.identifier.pmid | 22752320 | |
dc.identifier.doi | 10.1259/dmfr/20920453 | |
dc.identifier.uri | http://hdl.handle.net/10541/263679 | |
dc.description.abstract | To characterize the dose distribution for a range of cone beam CT (CBCT) units, investigating different field of view sizes, central and off-axis geometries, full or partial rotations of the X-ray tube and different clinically applied beam qualities. The implications of the dose distributions on the definition and practicality of a CBCT dose index were assessed. | |
dc.language.iso | en | en |
dc.rights | Archived with thanks to Dento maxillo facial radiology | en_GB |
dc.subject.mesh | Cone-Beam Computed Tomography | |
dc.subject.mesh | Equipment Design | |
dc.subject.mesh | Film Dosimetry | |
dc.subject.mesh | Humans | |
dc.subject.mesh | Phantoms, Imaging | |
dc.subject.mesh | Polymethyl Methacrylate | |
dc.subject.mesh | Quality Control | |
dc.subject.mesh | Radiation Dosage | |
dc.subject.mesh | Radiation Monitoring | |
dc.subject.mesh | Radiography, Dental | |
dc.subject.mesh | Scattering, Radiation | |
dc.subject.mesh | Thermoluminescent Dosimetry | |
dc.subject.mesh | Water | |
dc.title | Dose distribution for dental cone beam CT and its implication for defining a dose index. | en |
dc.type | Article | en |
dc.contributor.department | Department of Oral Health Sciences, Leuven, Belgium. | en_GB |
dc.identifier.journal | Dentomaxillofacial Radiology | en_GB |
html.description.abstract | To characterize the dose distribution for a range of cone beam CT (CBCT) units, investigating different field of view sizes, central and off-axis geometries, full or partial rotations of the X-ray tube and different clinically applied beam qualities. The implications of the dose distributions on the definition and practicality of a CBCT dose index were assessed. |