Affiliation
Regional Department of Medical Physics and Bioengineering, Christie Hospital, Manchester, England.Issue Date
1987
Metadata
Show full item recordAbstract
Lithium fluoride (LiF) thermoluminescent dosimeters (TLD) have been employed to measure the radiation dose distribution within a phantom and the central axis dose in air. Results are presented for seven (four EMI CT1010, one EMI CT5005, one EMI CT7070 and one Siemens DR2) different machines. Organ doses for four different computed tomography (CT) investigations (head, lung, liver and pelvis) have been estimated from previously published tables and the central axis dose in air in 3 CT units. These estimated organ doses were used in turn to deduce the somatic and genetic risks for the four types of CT examination. These calculations indicate that for an 'average' male patient undergoing a CT examination corresponding to the average conditions encountered, the total somatic risks are 3.15 10(-4), 1.98 10(-4), 2.31 10(-4), and 1.38 10(-4) for head, lung, liver and pelvis scans, respectively. The corresponding figures for female patients are 3.39 10(-4), 3.95 10(-4), 2.73 10(-4), and 1.60 10(-4). The risk from head scanning is approximately 250 times that of a dental pantomograph. Somatic and genetic risks will be approximately twice as high for contrast examinations.Citation
Radiation dose and somatic risk from computed tomography., 28 (4):483-8 Acta RadiolJournal
Acta RadiologicaPubMed ID
2958068Type
ArticleLanguage
enISSN
0284-1851Collections
Related articles
- Radiation dose and risk from fluoroscopically guided percutaneous transhepatic biliary procedures.
- Authors: Stratakis J, Damilakis J, Hatzidakis A, Perisinakis K, Gourtsoyiannis N
- Issue date: 2006 Jan
- [Estimation of organ dose during radiological and computer tomographic examinations with calculations of the somatically significant doses (author's transl)].
- Authors: Ewen K, Steiner H, Jungblut R, Günther D, Schoppe WD
- Issue date: 1980 Oct
- A Monte Carlo based method to estimate radiation dose from multidetector CT (MDCT): cylindrical and anthropomorphic phantoms.
- Authors: DeMarco JJ, Cagnon CH, Cody DD, Stevens DM, McCollough CH, O'Daniel J, McNitt-Gray MF
- Issue date: 2005 Sep 7
- [Radiation-exposure studies of CT diagnosis with the standard and spiral technics].
- Authors: Langkowski JH, Pogoda P, Hess A
- Issue date: 1994 Jul
- Current status of patient radiation doses from computed tomography examinations in Tanzania.
- Authors: Ngaile JE, Msaki P, Kazema R
- Issue date: 2006