Full 3D position reconstruction of a radioactive source based on a novel hyperbolic geometrical algorithm
Panaino, C. M. V. ; Mackay, Ranald I ; Sotiropoulos, M. ; Kirkby, Karen J ; Taylor, Michael J
Panaino, C. M. V.
Mackay, Ranald I
Sotiropoulos, M.
Kirkby, Karen J
Taylor, Michael J
Citations
Altmetric:
Abstract
A new method to locate, with millimetre uncertainty, in 3D, an x-ray source emitting multiple X-rays in a cascade, employing conventional LaBr3(Ce) scintillation detectors, has been developed. Using 16 detectors in a symmetrical configuration the detector energy and time signals, resulting from the X-ray interactions, are fed into a new source position reconstruction algorithm. The Monte-Carlo based Geant4 framework has been used to simulate the detector array and a 60Co source located at two positions within the spectrometer central volume. For a source located at (0,0,0) the algorithm reports X, Y, Z values of -0.3 ± 2.5, -0.4 ± 2.4, and -0.6 ± 2.5 mm, respectively. For a source located at (20,20,20) mm, with respect to the array centre, the algorithm reports X, Y, Z values of 20.2 ± 1.0, 20.2 ± 0.9, and 20.1 ± 1.2 mm. The resulting precision of the reconstruction means that this technique could find application in a number of areas including nuclear medicine, national security, radioactive waste assay and proton beam therapy.
Keywords: Gamma-ray spectroscopy; Reconstruction algorithm; Source emission position reconstruction.
Description
Date
2020
Publisher
Collections
Files
Loading...
From UNPAYWALL
Adobe PDF, 2.79 MB
Keywords
Type
Article
Citation
Panaino CMV, Mackay RI, Sotiropoulos M, Kirkby KJ, Taylor MJ. Full 3D position reconstruction of a radioactive source based on a novel hyperbolic geometrical algorithm. Comput Phys Commun. 2020;252:107131.