Assessing the inter and intra-fraction changes utilising MR Linac for cervical cancer
Alshamrani, A. ; Chuter, Robert ; Nelder, Claire L ; Choudhury, Ananya ; Barraclough, Lisa H ; Eccles, Cynthia L ; Aznar, Marianne Camille ; Hoskin, Peter J
Alshamrani, A.
Chuter, Robert
Nelder, Claire L
Choudhury, Ananya
Barraclough, Lisa H
Eccles, Cynthia L
Aznar, Marianne Camille
Hoskin, Peter J
Citations
Altmetric:
Abstract
Purpose or Objective
Online adaptive radiotherapy based on MR Linac (MRL) can potentially address inter and intra-fractional changes using
images acquired daily during treatment. We report the intra- and inter-fractional movement for the first 25 fraction cervical
cancer treatment delivered on the MRL in the UK.
Materials and Methods
Following institutional approval, a woman with stage FIGO-IIB node-negative cervical cancer was consented to the
MOMENTUM trial (NCT04075305) for 25 fractions of radiotherapy. Treatment was prescribed according to the EMBRACE II
guidelines using adapt-to-shape (ATS) workflow. A comfortably full bladder was required for treatment.
Three MRIs were acquired at each fraction: a setup image ( used for plan adaptation), a verification (verif) image (acquired
immediately prior to ‘beam on’) and a post-treatment (post-treat) image (to assess intra-fractional changes). A single
observer contoured the bladder and CTV on all available verif and post-treat MRIs. Intra-fraction changes in bladder and
CTV were determined by measuring the volume differences between the setup and verif images and the volume differences
between the verif and post-treat images. The inter-fraction changes in CTV and bladder volumes were determined by
measuring the difference between the verif images from each fraction and the MRI used for initial treatment
planning. Intra-fractional changes were reported using Dice coefficients (DC) and inter-fraction changes reported using
volume changes in cm3. Statistics have been presented as descriptive statistics and boxplots.
Results
22 fractions and 67 MR images were available for review. The mean treatment time was 68 min (range 51–82 min) from
the time the patient entered the room until the patient left. The median time from the setup image to verif image was
43.5 min (range 31-72 min). The median time from verif image to post-treat image was 13 min (range 4-40 min). The median
intra-fraction DC for the CTV was 0.62 (range 0.40-0.80) and 0.90 (range 0.57-0.95) for the setup to verif and verif to post treat images respectively. The median intra-fractional DC in bladder volume from setup to verif images and from verif to post-treat images were 0.67 (range 0.44-0.83) and 0.90 (range 0.54-0.97) respectively. Figure 1 shows the intra-fraction
variation from setup to verif and verif to post treat images for both bladder and CTV. CTV volume changes were greatest
in the adaption time, and smaller during radiation delivery. For inter-fraction movement, median volume for CTV and
bladder were 7.41cm3 (range -4.30-24.92) and -133cm3 (range-382.63-57), respectively. No consistent bladder or CTV
volume patterns were observe. Conclusion
Drinking protocols were not optimal to keep bladder volume constant. The low DC value justifies the need for an ATS
workflow. Real-time adaptation based on MRL addressed inter and intra-fraction CTV and bladder changes by allowing an
online correction that improves delineation and treats on the most updated plan.