Quantitative comparison of 3D and 2.5D gamma analysis: introducing gamma angle histograms.
Al Sa'd, M ; Graham, J ; Liney, G P ; Moore, Christopher J
Al Sa'd, M
Graham, J
Liney, G P
Moore, Christopher J
Citations
Altmetric:
Abstract
Comparison of dose distributions using the 3D gamma method is anticipated to provide better indicators for the quality assurance process than the 2.5D (stacked 2D slice-by-slice) gamma calculation, especially for advanced radiotherapy technologies. This study compares the accuracy of the 3D and 2.5D gamma calculation methods. 3D and 2.5D gamma calculations were carried out on four reference/evaluation 3D dose sample pairs. A number of analysis methods were used, including average gamma and gamma volume histograms. We introduce the concept of gamma-angle histograms. Noise sensitivity tests were also performed using two different noise models. The advantage of the 3D gamma method showed up as a higher proportion of points passing the tolerance criteria of 3% dose difference and 3 mm distance-to-agreement (DTA), with considerably lower average gamma values, a lower influence of the DTA criterion, and a higher noise tolerance. The 3D gamma approach is more reliable than the 2.5D approach in terms of providing comprehensive quantitative results, which are needed in quality assurance procedures for advanced radiotherapy methods.
Description
Date
2013-04-21
Publisher
Collections
Keywords
Type
Article
Citation
Quantitative comparison of 3D and 2.5D gamma analysis: introducing gamma angle histograms. 2013, 58 (8):2597-608 Phys Med Biol