Cone beam CT with zonal filters for simultaneous dose reduction, improved target contrast and automated set-up in radiotherapy.
Moore, Christopher J ; Marchant, Thomas E ; Amer, Ali M
Moore, Christopher J
Marchant, Thomas E
Amer, Ali M
Citations
Altmetric:
Abstract
Cone beam CT (CBCT) using a zonal filter is introduced. The aims are reduced concomitant imaging dose to the patient, simultaneous control of body scatter for improved image quality in the tumour target zone and preserved set-up detail for radiotherapy. Aluminium transmission diaphragms added to the CBCT x-ray tube of the Elekta Synergytrade mark linear accelerator produced an unattenuated beam for a central "target zone" and a partially attenuated beam for an outer "set-up zone". Imaging doses and contrast noise ratios (CNR) were measured in a test phantom for transmission diaphragms 12 and 24 mm thick, for 5 and 10 cm long target zones. The effect on automatic registration of zonal CBCT to conventional CT was assessed relative to full-field and lead-collimated images of an anthropomorphic phantom. Doses along the axis of rotation were reduced by up to 50% in both target and set-up zones, and weighted dose (two thirds surface dose plus one third central dose) was reduced by 10-20% for a 10 cm long target zone. CNR increased by up to 15% in zonally filtered CBCT images compared to full-field images. Automatic image registration remained as robust as that with full-field images and was superior to CBCT coned down using lead-collimation. Zonal CBCT significantly reduces imaging dose and is expected to benefit radiotherapy through improved target contrast, required to assess target coverage, and wide-field edge detail, needed for robust automatic measurement of patient set-up error.
Description
Date
2006-05-07
Publisher
Collections
Keywords
Type
Article
Citation
Cone beam CT with zonal filters for simultaneous dose reduction, improved target contrast and automated set-up in radiotherapy. 2006, 51 (9):2191-204 Phys Med Biol