Large area CMOS active pixel sensor x-ray imager for digital breast tomosynthesis: Analysis, modeling, and characterization.
Zhao, C ; Kanicki, J ; Konstantinidis, Anastasios C ; Patel, T
Zhao, C
Kanicki, J
Konstantinidis, Anastasios C
Patel, T
Citations
Altmetric:
Abstract
Large area x-ray imagers based on complementary metal-oxide-semiconductor (CMOS) active pixel sensor (APS) technology have been proposed for various medical imaging applications including digital breast tomosynthesis (DBT). The low electronic noise (50-300 e(-)) of CMOS APS x-ray imagers provides a possible route to shrink the pixel pitch to smaller than 75 μm for microcalcification detection and possible reduction of the DBT mean glandular dose (MGD).
Description
Date
2015-11
Publisher
Collections
Files
Loading...
From UNPAYWALL
Adobe PDF, 1.13 MB
Keywords
Type
Article
Citation
Large area CMOS active pixel sensor x-ray imager for digital breast tomosynthesis: Analysis, modeling, and characterization. 2015, 42 (11):6294 Med Phys