Abstract:
A technique for deriving the reference air kerma rate for low activity brachytherapy sources that is traceable to the therapy level standard at the National Physical Laboratory is presented. Correction factors have been generated to account for the finite source and detector size. The air kerma rate calibration of the secondary standard for caesium quality has been derived by polynomial curve fitting. The reference air kerma rates for several caesium tubes and needles have been determined and the results compared with the manufacturers' source test reports. For all the source types used the agreement between methods was within 2%.
Affiliation:
North Western Medical Physics Department, Christie Hospital, Manchester, UK.
Citation:
Calibration of low activity caesium tubes and needles traceable to the therapy level standard. 1994, 67 (794):194-9 Br J Radiol
Journal:
The British Journal of Radiology
Issue Date:
Feb-1994
URI:
http://hdl.handle.net/10541/97034
DOI:
10.1259/0007-1285-67-794-194
PubMed ID:
8130983
Type:
Article
Language:
en
ISSN:
0007-1285
Appears in Collections:
All Paterson Institute for Cancer Research

Full metadata record

DC FieldValue Language
dc.contributor.authorDeshpande, N Aen
dc.contributor.authorWilkinson, John Men
dc.date.accessioned2010-04-21T11:26:09Z-
dc.date.available2010-04-21T11:26:09Z-
dc.date.issued1994-02-
dc.identifier.citationCalibration of low activity caesium tubes and needles traceable to the therapy level standard. 1994, 67 (794):194-9 Br J Radiolen
dc.identifier.issn0007-1285-
dc.identifier.pmid8130983-
dc.identifier.doi10.1259/0007-1285-67-794-194-
dc.identifier.urihttp://hdl.handle.net/10541/97034-
dc.description.abstractA technique for deriving the reference air kerma rate for low activity brachytherapy sources that is traceable to the therapy level standard at the National Physical Laboratory is presented. Correction factors have been generated to account for the finite source and detector size. The air kerma rate calibration of the secondary standard for caesium quality has been derived by polynomial curve fitting. The reference air kerma rates for several caesium tubes and needles have been determined and the results compared with the manufacturers' source test reports. For all the source types used the agreement between methods was within 2%.en
dc.language.isoenen
dc.subject.meshBrachytherapy-
dc.subject.meshCalibration-
dc.subject.meshCesium Radioisotopes-
dc.subject.meshHumans-
dc.subject.meshNeedles-
dc.subject.meshRadiometry-
dc.subject.meshRadiotherapy Dosage-
dc.subject.meshReference Standards-
dc.titleCalibration of low activity caesium tubes and needles traceable to the therapy level standard.en
dc.typeArticleen
dc.contributor.departmentNorth Western Medical Physics Department, Christie Hospital, Manchester, UK.en
dc.identifier.journalThe British Journal of Radiologyen
All Items in Christie are protected by copyright, with all rights reserved, unless otherwise indicated.