Title:
Quality control aspects of the Philips multileaf collimator.
Authors:
Hounsell, Alan R; Jordan, Thomas J
Abstract:
BACKGROUND AND PURPOSE: Linear accelerators equipped with multileaf collimators (MLCs) are becoming more common and are widely available from most commercial manufacturers. There is a need to ensure they retain their commissioning specification using a preventative maintenance and quality control (QC) programme. This paper considers the design criteria of the Philips MLC which are important to the production of a comprehensive quality control programme. MATERIALS AND METHODS: The specific QC problems related to MLCs are identified as the positional accuracy of the leaves and their relationship to the back-up collimators, leakage considerations, the relationship of X-ray to light field and the influence of gravity on the positioning and leakage characteristics of the leaves. These problems are considered in relation to the general design considerations of the MLC, and methods of performing routine quality control checks are discussed. RESULTS AND CONCLUSIONS: The introduction of MLCs into clinical use results in new QC procedures being developed but it can be concluded that for the Philips MLC only an extra 30 min of QC time is needed per month and that its use has added little to the general down-time of this department.
Affiliation:
North Western Medical Physics, Christie Hospital NHS Trust, Manchester, UK.
Citation:
Quality control aspects of the Philips multileaf collimator. 1997, 45 (3):225-33 Radiother Oncol
Journal:
Radiotherapy and Oncology
Issue Date:
Dec-1997
URI:
http://hdl.handle.net/10541/94681
DOI:
10.1016/S0167-8140(97)00100-X
PubMed ID:
9426116
Type:
Article
Language:
en
ISSN:
0167-8140
Appears in Collections:
All Christie Publications

Full metadata record

DC FieldValue Language
dc.contributor.authorHounsell, Alan Ren
dc.contributor.authorJordan, Thomas Jen
dc.date.accessioned2010-03-23T12:56:59Z-
dc.date.available2010-03-23T12:56:59Z-
dc.date.issued1997-12-
dc.identifier.citationQuality control aspects of the Philips multileaf collimator. 1997, 45 (3):225-33 Radiother Oncolen
dc.identifier.issn0167-8140-
dc.identifier.pmid9426116-
dc.identifier.doi10.1016/S0167-8140(97)00100-X-
dc.identifier.urihttp://hdl.handle.net/10541/94681-
dc.description.abstractBACKGROUND AND PURPOSE: Linear accelerators equipped with multileaf collimators (MLCs) are becoming more common and are widely available from most commercial manufacturers. There is a need to ensure they retain their commissioning specification using a preventative maintenance and quality control (QC) programme. This paper considers the design criteria of the Philips MLC which are important to the production of a comprehensive quality control programme. MATERIALS AND METHODS: The specific QC problems related to MLCs are identified as the positional accuracy of the leaves and their relationship to the back-up collimators, leakage considerations, the relationship of X-ray to light field and the influence of gravity on the positioning and leakage characteristics of the leaves. These problems are considered in relation to the general design considerations of the MLC, and methods of performing routine quality control checks are discussed. RESULTS AND CONCLUSIONS: The introduction of MLCs into clinical use results in new QC procedures being developed but it can be concluded that for the Philips MLC only an extra 30 min of QC time is needed per month and that its use has added little to the general down-time of this department.en
dc.language.isoenen
dc.subject.meshCalibration-
dc.subject.meshParticle Accelerators-
dc.subject.meshQuality Control-
dc.subject.meshTechnology, Radiologic-
dc.titleQuality control aspects of the Philips multileaf collimator.en
dc.typeArticleen
dc.contributor.departmentNorth Western Medical Physics, Christie Hospital NHS Trust, Manchester, UK.en
dc.identifier.journalRadiotherapy and Oncologyen
All Items in Christie are protected by copyright, with all rights reserved, unless otherwise indicated.