Requirements for leaf position accuracy for dynamic multileaf collimation.

2.50
Hdl Handle:
http://hdl.handle.net/10541/86387
Title:
Requirements for leaf position accuracy for dynamic multileaf collimation.
Authors:
Budgell, Geoff J; Mott, Judith; Williams, Peter C; Brown, K J
Abstract:
Intensity modulated radiation therapy can be achieved by driving the leaves of a multileaf collimator (MLC) across an x-ray therapy beam. Algorithms to generate the required leaf trajectories assume that the leaf positions are exactly known to the MLC controller. In practice, leaf positions depend upon calibration accuracy and stability and may vary within set tolerances. The purpose of this study was to determine the effects of potential leaf position inaccuracies on intensity modulated beams. Equations are derived which quantify the absolute error in delivered monitor units given a known error in leaf position. The equations have been verified by ionization chamber measurements in dynamically delivered flat fields, comparing deliveries in which known displacements have been applied to the defined leaf positions with deliveries without displacements applied. The equations are then applied to two clinical intensity modulations: an inverse planned prostate field and a breast compensating field. It is shown that leaf position accuracy is more critical for a highly modulated low-dose intensity profile than a moderately modulated high-dose intensity profile. Suggestions are given regarding the implications for quality control of dynamic MLC treatments.
Affiliation:
North Western Medical Physics, Christie Hospital NHS Trust, Manchester, UK. Geoff.Budgell@physics.cr.man.ac.uk
Citation:
Requirements for leaf position accuracy for dynamic multileaf collimation. 2000, 45 (5):1211-27 Phys Med Biol
Journal:
Physics in Medicine and Biology
Issue Date:
May-2000
URI:
http://hdl.handle.net/10541/86387
DOI:
10.1088/0031-9155/45/5/310
PubMed ID:
10843101
Type:
Article
Language:
en
ISSN:
0031-9155
Appears in Collections:
All Christie Publications

Full metadata record

DC FieldValue Language
dc.contributor.authorBudgell, Geoff Jen
dc.contributor.authorMott, Judithen
dc.contributor.authorWilliams, Peter Cen
dc.contributor.authorBrown, K Jen
dc.date.accessioned2009-11-18T12:24:22Z-
dc.date.available2009-11-18T12:24:22Z-
dc.date.issued2000-05-
dc.identifier.citationRequirements for leaf position accuracy for dynamic multileaf collimation. 2000, 45 (5):1211-27 Phys Med Biolen
dc.identifier.issn0031-9155-
dc.identifier.pmid10843101-
dc.identifier.doi10.1088/0031-9155/45/5/310-
dc.identifier.urihttp://hdl.handle.net/10541/86387-
dc.description.abstractIntensity modulated radiation therapy can be achieved by driving the leaves of a multileaf collimator (MLC) across an x-ray therapy beam. Algorithms to generate the required leaf trajectories assume that the leaf positions are exactly known to the MLC controller. In practice, leaf positions depend upon calibration accuracy and stability and may vary within set tolerances. The purpose of this study was to determine the effects of potential leaf position inaccuracies on intensity modulated beams. Equations are derived which quantify the absolute error in delivered monitor units given a known error in leaf position. The equations have been verified by ionization chamber measurements in dynamically delivered flat fields, comparing deliveries in which known displacements have been applied to the defined leaf positions with deliveries without displacements applied. The equations are then applied to two clinical intensity modulations: an inverse planned prostate field and a breast compensating field. It is shown that leaf position accuracy is more critical for a highly modulated low-dose intensity profile than a moderately modulated high-dose intensity profile. Suggestions are given regarding the implications for quality control of dynamic MLC treatments.en
dc.language.isoenen
dc.subjectBreast Canceren
dc.subjectProstatic Canceren
dc.subject.meshAlgorithms-
dc.subject.meshBiophysical Phenomena-
dc.subject.meshBiophysics-
dc.subject.meshBreast Neoplasms-
dc.subject.meshFemale-
dc.subject.meshHumans-
dc.subject.meshMale-
dc.subject.meshModels, Theoretical-
dc.subject.meshProstatic Neoplasms-
dc.subject.meshRadiotherapy, Conformal-
dc.titleRequirements for leaf position accuracy for dynamic multileaf collimation.en
dc.typeArticleen
dc.contributor.departmentNorth Western Medical Physics, Christie Hospital NHS Trust, Manchester, UK. Geoff.Budgell@physics.cr.man.ac.uken
dc.identifier.journalPhysics in Medicine and Biologyen

Related articles on PubMed

All Items in Christie are protected by copyright, with all rights reserved, unless otherwise indicated.